Comparison between noninvasive hemoglobin(SpHb) monitoring and venous hemoglobin measurement in emergency department patients |
Huang Wei-chang, Li Qi-ling, Xu Xue-lian, Liang Zhi-hao, Liu Bin |
Department of Emergency, Zhujiang Hospital, Southern Medical University, Guangzhou 510280, China |
|
|
Abstract Objective To explore the accuracy of noninvasive hemoglobin(SpHb) monitoring in emergency patients, and to evaluate its application value in Emergency Department. Methods 300 patients in Emergency Department of Zhujiang Hospital, Southern Medical University, from October 2017 to January 2018, were enrolled finally. Noninvasive hemoglobin monitoring by spectrophotometry was reported by NT1G, while the venous blood samples were collected and true hemoglobin level (tHb) was measured by Sysmex XE-5000 automatic hematology analyzer. The precision which SpHb compared with tHb was determined by using both Pearson correlation and Bland-Altman analysis. Results Three hundred hemoglobin measurements were collected in this study. Excellent correlation was reported between both measures with Pearson correlation coefficient of 0.862 (P<0.001), and regression equation of y=-0.239+1.071x. Bland-Altman analysis showed that the average difference between SpHb and tHb was -0.59 g/dL (bias) with a standard deviation of the mean bias of 1.09 g/dL, and the 95% CI of agreement was -2.73~1.55 g/dL. The bias between SpHb and tHb was exceeding the clinically acceptable range, which meaned that the consistency was not enough. The chi-square test of the fourfold table showed that the diagnosis of blood transfusion with SpHb was of high sensitivity, high specificity, and high diagnostic accuracy. Conclusion There is a significant correlation between SpHb and tHb in emergency department patients. SpHb measurement can be used for predicting the tHb, and monitoring the changes of hemoglobin in emergency departments. SpHb plays an important role in guiding procedures for blood transfusion decision-making. However, SpHb monitoring could not replace the venous hemoglobin measurement because of the low consistency. The accuracy of SpHb monitoring needs to be further improved.
|
|
Corresponding Authors:
Liu Bin, E-mail: nysylb@163.com
|
|
|
|
[1]Seguin P, Kleiber A, Chanavaz C, et al. Determination of capillary hemoglobin levels using the HemoCue system in intensive care patients[J]. J Crit Care, 2011, 26(4): 423-427.
[2]Macknet MR, Allard M, Applegate RL 2nd, et al. The accuracy of noninvasive and continuous total hemoglobin measurement by pulse CO-oximetry in human subjects undergoing hemodilution[J]. Anesth Analg, 2010, 111(6):1424-1426.
[3]Awada WN, Mohmoued MF, Radwan TM, et al. Continuous and noninvasive hemoglobin monitoring reduces red blood cell transfusion during neurosurgery: a prospective cohort study[J]. J Clin Monit Comput, 2015, 29(6): 733-740.
[4]Al-Khabori M, Al-Riyami AZ, Al-Farsi K, et al. Validation of a non-invasive pulse CO-oximetry based hemoglobin estimation in normal blood donors[J]. Transfus Apher Sci, 2014, 50(1): 95-98.
[5]Yamada H, Saeki M, Ito J, et al. The relative trending accuracy of noninvasive continuous hemoglobin monitoring during hemodialysis in critically ill patients[J]. J Clin Monit Comput, 2015, 29(1):107-112.
[6]Stoesz M, Wood K, Clark W, et al. Utility of noninvasive transcutaneous measurement of postoperative hemoglobin in total joint arthroplasty patients[J]. J Arthroplasty, 2014, 29(11): 2084-2086.
[7]Xu T, Yang T, Kim JB, et al. Evaluation of noninvasive hemoglobin monitoring in surgical critical care patients[J]. Crit Care Med, 2016, 44(6): 344-352.
[8]陈卉.Bland-Altman分析在临床测量方法一致性评价中的应用[J].中国卫生统计, 2007, 24(3):308-309,315.
[9]岑宪铭.限制性红细胞输血策略的依据及输血阈值[J]. 临床输血与检验, 2016, 18(1):90-94.
[10]Gamal M, Abdelhamid B, Zakaria D, et al. Evaluation of noninvasive hemoglobin monitoring in trauma patients with low hemoglobin levels[J].Shock, 2018, 49(2):150-153.
[11]Nguyen BV, Vincent JL, Nowak E, et al. The accuracy of noninvasive hemoglobin measurement by multiwave length pulse oximetry after cardiac surgery[J].Anesth Analg, 2011, 113(5):1052-1057.
[12]Zeng R, Svensen CH, Li H, et al. Can noninvasive hemoglobin measurement reduce the need for preoperative venipuncture in pediatric outpatient surgery[J].Paediatr Anaesth, 2017, 27(11):1131-1135.
[13]徐勇, 罗佛全, 赵为禄, 等. 无创血红蛋白监测在脊柱融合术中的应用研究[J].临床麻醉学杂志, 2015, 31(3):217-220.
[14]Vos JJ, Kalmar AF, Struys MM, et al. Accuracy of non-invasive measurement of haemoglobin concentration by pulse co-oximetry during steady-state and dynamic conditions in liver surgery[J]. Br J Anaesth, 2012, 109(4): 522-528.
[15]Mntele W, Deniz E. UV-VIS absorption spectroscopy: Lambert-Beer reloaded[J]. Spectrochim Acta A Mol Biomol Spectrosc, 2017, 173(2): 965-968.
[16]Sjstrand F, Rodhe P, Berglund E, et al. The use of a noninvasive hemoglobin monitor for volume kinetic analysis in an emergency room setting[J]. Anesth Analg, 2013, 116(2): 337-342. |
|
|
|